June 12, 2008
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Ventana Medical Systems Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS, N750-BMKLT-FS, N750-DIS-FS, N750-DISXT-FS. Product Usage: Usage: The Benchmark product lines are intended to automatically stain histological or cytological specimens on microscope slides with specific immunohistochemistry or in situ hybridization reagents for in vitro diagnostic use. The Discovery Systems are research use only instruments with similar functionality.

Z-2118-2013
Recall number
Z-2118-2013
Initiated
June 12, 2008
Classification
Class II
Status
Terminated
Quantity
1217 units total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the heating element could separate from the underside of the heater cap and cause a non-staining or inconsistent staining event. Additionally, the system may not detect this anomaly and therefore it may not generate an error message for the slide position.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the heating element could separate from the underside of the heater cap and cause a non-staining or inconsistent staining event. Additionally, the system may not detect this anomaly and therefore it may not generate an error message for the slide position.

Code information

BenchMark XT Instruments-S/N: 711869-712465; BenchMark LT instruments-S/N: 610321-610345. Discovery XT Instruments-S/N: 610321-610345.

Distribution pattern