epoc Blood Analysis BGEM Test Cards for use with the epoc(R) Blood Analysis System Product Usage: Usage: epoc Blood Analysis BGEM Test Cards are used with the epoc(R) Blood Analysis System which is a POC testing system
Customer reported problem while performing incoming QC of newly received lot of test cards. Liquid quality control results were found to be out of range. Further investigation by Epocal determined that a portion of the test card lot (approximately 3000 cards) could potentially report low glucose results.
These labels are deterministic app interpretations, not FDA categories.
Customer reported problem while performing incoming QC of newly received lot of test cards. Liquid quality control results were found to be out of range. Further investigation by Epocal determined that a portion of the test card lot (approximately 3000 cards) could potentially report low glucose results.