Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 61946

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 21, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Natus Medical Incorporated

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Olympic Cool-Cap System is intended to provide treatment for neonatal hypoxic-ischemic encephalopathy (HIE). The Cool-Cap is indicated for use in full term infants (more than 36 week gestational age) with clinical evidence of moderate to severe HIE. The device provides selective head cooling with mild systemic hypothermia to prevent or reduce the severity of neurologic injury associated with HIE. It is a selective head cooling system for treatment of neonatal hypoxic-ischemic encephalopathy (HIE) in infants. A cooling water cap is placed on the infant's head and the body is warmed using radiant warmers. The Cool-Cap provides 72 hours of monitored cooling. The device is labeled in part: "***OLYMPIC COOL-CAP Cooling Unit***Model No.***Made in USA***Software in this product is Natus Medical Incorporated***"

Z-1843-2012
Recall number
Z-1843-2012
Initiated
May 21, 2012
Classification
Class II
Status
Terminated
Quantity
181 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Olympic Cool-Cap Cooling Module's power supply failed during treatment. The power supply failure can happen without prior indication and results in the unavailability of the system to begin or continue cooling treatment.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Olympic Cool-Cap Cooling Module's power supply failed during treatment. The power supply failure can happen without prior indication and results in the unavailability of the system to begin or continue cooling treatment.

Code information

Cooling Serial Numbers: 1008; 1009; 1011; 1012; 1014; 1016; 1017; 1018; 1020; 1021; 1022; 1023; 1024; 1025; 1027; 1028; 1030; 1031; 1034; 1035; 1036; 1038; 1044; 1045; 1048; 1050; 1051; 1052; 1053; 1054; 1055; 1056; 1059; 1060; 1062; 1063; 1064; 1065; 1066; 1070; 1071; 1072; 1073; 1074; 1075; 1076; 1077; 1080; 1081; 1082; 1083; 1085; 1086; 1087; 1088; 1090; 1091; 1092; 1093; 1094; 1095; 1096; 1097; 1101; 1116; 1117; 1118; 1119; 1120; 1122; 1124; 1125; 1126; 1127; 1128; 1129; 1130; 1131; 1132; 1136; 1137; 1138; 1139; 1141; 7002; 7003; 7004; 7005; 7007; 7008; 7009; 7010; 7011; 7013; 7014; 7016; 7017; 7018; 7019; 7020; 7021; 7022; 7023; 7024; 7025; 7026; 7027; 7028; 7030; 7031; 7032; 7033; 7034; 7036; 7037; 7038; 7039; 7040; 7041; 7042; 7043; 7044; 7045; 7046; 7047; 7048; 7049; 7051; 7052; 7053; 7054; 7055; 7250; 7251; 7252; 7253; 7255; 7256; 7257; 7258; 7259; 7260; 7261; 7262; 7263; 7264; 7265; 7266; 7267; 7268; 7269; 7270; 7271; 7272; 7273; 7274; 7275; 7276; 7277; 7278; 7279; 7280; 7281; 7282; 7283; 7284; 7285; 7286; 7288; 7289; 7290; 7291; 7292; 7293; 7294; 7363; 7364; 7365; 7366; 7372; and 7374.

Distribution pattern

Worldwide Distribution -- USA (nationwide) including the states of : AK; AR; AZ; CA; CO; DE; FL; GA; HI; ID; IL; IN; KY; MA; MI; MN; MS; ND; NE; NJ; NY; OH; PA; SC; SD; TN; TX; and WA. and the countries of : Argentina; Australia; Austria; Canada; Egypt; Germany; Hong Kong; Indonesia; Iran; Italy; Kuwait; Poland; Russia; Saudi Arabia; Singapore; Slovenia; South Africa; South Korea; Spain; Taiwan; Thailand; Turkey; United Arab Emirates; and United Kingdom.