Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 62007

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 25, 2012
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
CareFusion 303, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Alaris PC unit model 8015 Product Usage: The device is labeled for prescription use only.

Z-2076-2012
Recall number
Z-2076-2012
Initiated
May 25, 2012
Classification
Class I
Status
Terminated
Recalling firm
CareFusion 303, Inc.
Quantity
112,912 units total (106,888 units in US)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The recall was initiated because the Alaris PC unit model 8015 has a component on the PC unit power supply board that is causing an error code(120.4630) "System Error" or "Missing Battery Error" at start up. The error code is accompanied by both an audible alarm and visual error messages on the PC unit screen.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The recall was initiated because the Alaris PC unit model 8015 has a component on the PC unit power supply board that is causing an error code(120.4630) "System Error" or "Missing Battery Error" at start up. The error code is accompanied by both an audible alarm and visual error messages on the PC unit screen.

Code information

not available

Distribution pattern

US Nationwide Distribution