openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
JUGGERKNOT IN¿GUIDE PUNCH, 2.9MM. The JuggerKnot 2.9mm In-Guide Punch is intended to be used to prepare a 2.9mm hole into bone in which the JuggerKnot 2.9mm Suture Anchor is to be placed. The JuggerKnot 2.9mm Suture Anchor is intended to be used to attach soft tissue to bone.
Investigation found that the current design of the product allows the press-fit of the knob/shaft interface to be overcome by the impaction force of a mallet, causing the knob to potentially unfasten and slide down the shaft.
These labels are deterministic app interpretations, not FDA categories.
Investigation found that the current design of the product allows the press-fit of the knob/shaft interface to be overcome by the impaction force of a mallet, causing the knob to potentially unfasten and slide down the shaft.
Code information
REF. 912064 LOT 121040, 497540, 839530
Distribution pattern
Nationwide Distribution-including the states of WA, IN, AK, PA, GA, SD, KS, NY, VA, AR, AL, OH, WI, MA, and CA.