Event 62404

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 19, 2007
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
GE OEC Medical Systems, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

OEC 9900 Elite, OEC 9900 Elite MD Motorized C-arm System, OEC 9900 Elite NAV used as an aid to locate anatomical structures during open or percutaneous surgical procedures. The system is an aid to locate anatomical structures during open or percutaneous surgical procedures. It is indicated for use in medical conditions that may benefit from the use of stereotactic surgery technique. The system provides a reference to rigid anatomical structures such as sinus, skull, long bone, or vertebra, which are visible on medical images such as CT, MRI, or X-ray.

Z-2025-2012
Recall number
Z-2025-2012
Initiated
November 19, 2007
Classification
Class II
Status
Terminated
Quantity
726 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

Code information

Part numbers: 00-887200-01; 00-887200-03; 00-887201-01; 00-887201-03; 00-887202-01; 00-887202-03; 00-887204-03; 00-887207-03; 00-887206-03. Along with all associated serial numbers.

Distribution pattern

Worldwide Distribution.

device · product 2 of 3

OEC 9800, OEC FluoroTrak 9800 Plus, OEC 9800 Plus, OEC 9800 MD Motorized C-arm System designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The system includes features specifically designed for use in diagnostic and interventional cardiac imaging procedures, and is also intended for cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. It may be used for other imaging applications at the physician's discretion.

Z-2026-2012
Recall number
Z-2026-2012
Initiated
November 19, 2007
Classification
Class II
Status
Terminated
Quantity
11,100 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

Code information

Part numbers: 00-884909-01, 00-884777-01, 00-884911-01, 00-885553-01, 00-884913-01, 00-885554-01, 00-885555-01, 00-886289-01. All serial numbers.

Distribution pattern

Worldwide Distribution.

device · product 3 of 3

OEC 8800 Flexiview designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The OEC 8800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical applications may include cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The system may be used for other imaging applications at the physician's discretion.

Z-2027-2012
Recall number
Z-2027-2012
Initiated
November 19, 2007
Classification
Class II
Status
Terminated
Quantity
1349 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

Code information

Part numbers: OEC 8800 All serial numbers.

Distribution pattern

Worldwide Distribution.