Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 62451

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 17, 2009
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Kimberly-Clark Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

MicroCool Breathable Impervious Surgical Gown, (Large and X-Large), Sterile, Made in Honduras. The Kimberly Clark Micro COOL Surgical Gowns are sterile single use surgical apparel intended to be worn by heathcare professionals to help protect the patient and the healthcare worker from the transfer of microorganisms, body fluids and particulate matter.

Z-2017-2012
Recall number
Z-2017-2012
Initiated
September 17, 2009
Classification
Class II
Status
Terminated
Quantity
79,941 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints were received of gowns pilling or balling, linting and abrades, blood strikethrough and fibers falling off the gowns.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints were received of gowns pilling or balling, linting and abrades, blood strikethrough and fibers falling off the gowns.

Code information

Product Code: 92340 (Large), Lot #AH9103, AH9110, AH9139, AH9140, AG9154, AH9155 and AH9160; Product Code: 92342 (X-Large), Lot #AH9116, AH9142, AH9143, AH9145, AH9150, AH9151, AH9155, AH9157, AH9159, AH9160, AH9166, AH9168 and AH9076. Customers were notified to include Lot range AH9076xxx to AH9175xxx. On November 17, 2009 the recall was expanded to include the following: Product number 92340 (Large), Lot numbers: AH9072, AH9073, AH9074, AH9103, AH9140 and AH9146; Product number: 92342 (X-Large), Lot numbers: AH9073, AH9074, AH9112, AH9158, AH9160, AH9163, AH9165, AH9167 and AH9168; Product number: 92344 (XX-Large), Lot numbers: AH9074, AH9139, AH9152, AH9153, and AH9155; Product number: 92353 (X-Large Specialty), Lot numbers: AH9143, AH9146 and AH9160 and Product number: 92357 (XX-Large Specialty), Lot number range AH9072xxx to AH9175xxx.

Distribution pattern

Nationwide Distribution including AL, FL, GA, KY, LA, MS, NC, OH, SC, TN, and VA

device · product 2 of 2

MicroCool Breathable Impervious Gown with Secure Fit Surgical, (Large and X-Large), Made in Honduras. The Kimberly Clark Micro COOL Surgical Gowns are sterile single use surgical apparel intended to be worn by heathcare professionals to help protect the patient and the healthcare worker from the transfer of microorganisms, body fluids and particulate matter.

Z-2018-2012
Recall number
Z-2018-2012
Initiated
September 17, 2009
Classification
Class II
Status
Terminated
Quantity
Unavailable

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints were received of gowns pilling or balling, linting and abrades, blood strikethrough and fibers falling off the gowns.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints were received of gowns pilling or balling, linting and abrades, blood strikethrough and fibers falling off the gowns.

Code information

Product Code: 92354 (Secure-Fit Large), Lot #AH9141 and Product Code: 92355 (Secure Fit X-Large), Lot #AH9143. Customers were notified to include Lot range AH9076xxx to AH9175xxx. On November 17, 2009 the recall was expanded to include the following: Product number: 92354 (Large), Lot number range AH9072xxx to AH9175xxx; Product number: 92355 (X-Large), Lot number: AH9141 and Product number: 92358 (XX-Large), Lot number: AH9173.

Distribution pattern

Nationwide Distribution including AL, FL, GA, KY, LA, MS, NC, OH, SC, TN, and VA