Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 62516

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 15, 2011
Product types
Device
Classifications
Class III
Statuses
Terminated
Recalling firm wording
Sunquest Information Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Sunquest Laboratory : intended for use by professionals working in a clinical laboratory.

Z-0073-2013
Recall number
Z-0073-2013
Initiated
November 15, 2011
Classification
Class III
Status
Terminated
Quantity
127 sites

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Online calculations may not be performed as expected or may yield results that are different than expected when the calculation module is called from the Online Entry (OEM) application for a subtest whose subcolumn number on the instrument online string is divisible by 10.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Online calculations may not be performed as expected or may yield results that are different than expected when the calculation module is called from the Online Entry (OEM) application for a subtest whose subcolumn number on the instrument online string is divisible by 10.

Code information

Versions 6.4.0 and later

Distribution pattern

Nationwide distribution