Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 62557

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 14, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Steris Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

The Reliance 130 and 130L Cart Washer/Disinfector is a high capacity mechanical washer intended for use in the washing, low-level disinfecting, and drying of carts, utensils, beds and other miscellaneous reusable items used in the care of patients at healthcare facilities / Crated for shipping.

Z-2122-2012
Recall number
Z-2122-2012
Initiated
July 14, 2012
Classification
Class II
Status
Terminated
Recalling firm
Steris Corporation
Quantity
367

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
To ensure that users are properly operating the Reliance 130 and 130L Cart Washers, a door close confirmation button and door open delay will be installed on your cart washer.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

To ensure that users are properly operating the Reliance 130 and 130L Cart Washers, a door close confirmation button and door open delay will be installed on your cart washer.

Code information

The codes for the product are Julian date-coded. They follow the following pattern: AABBBCCDDD AA is a two digit code representing the manufacturing facility number BBB is a three digit code representing the day of the year that the device was manufactured CC is a two digit code representing the Julian year that the device was manufactured DDD is a three digit code representing the number of the number of the device manufactured that day. For example, a serial number of 3606499005 indicates that the device was manufactured in facility number 36, on the 64th day of the year of 1999, and the device was the 5th device manufactured that day. The serial number ranges for the devices being recalled are: RELIANCE¿ 130: 3606499005  3626709006 RELIANCE¿ 130L: 3629299006  3628809011

Distribution pattern

Worldwide Distribution -- USA (nationwide) including the states of : AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. and the countries of : Australia, South Korea, France, Italy, Lebanon, Russia, Saudi Arabia, Spain, UAE, United Kingdom and Canada.