Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 62594

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 02, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Stryker Howmedica Osteonics Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Stryker Orthopaedics Restoration Modular Cone Body Trials Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland. The Cone Body Trials simulate the respective Cone Body implants and mate with the stem trials to assist the surgeon in assessing the fit of the proposed Implants. The cone body trials is an assembly of the Body and 8 mm Hex Locking Bolt. The trial stem is attached to the trial Cone body via the distal thread of the Locking bolt.

Z-2487-2012
Recall number
Z-2487-2012
Initiated
April 02, 2012
Classification
Class II
Status
Terminated
Quantity
10, 983 units

App-derived interpretation

Foreign material or chemical contamination reason.foreign_material_chemical_contamination · v1.0.0
metal shaving

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Stryker Orthopaedics has become aware of a burr, or a metal shaving, observed in the internal threaded portion of the Cone Body Trial. The typical size of the shaving observed was 1-2 mm in length and 0.037 mm squared in cross section.

Code information

Catalog No. 6278-1-019 Mod Rev 19 mm std cone trial 6278-1-021 Mod Rev 21 mm std cone trial 6278-1-023 Mod Rev 23 mm std cone trial 6278-1-025 Mod Rev 25 mm std cone trial 6278-1-027 Mod Rev 27 mm std cone trial 6278-1-029 Mod Rev 29 mm std cone trial 6278-1-031 Mod Rev 31 mm std cone trial 6278-1-119 Mod Rev 19 mm + 10 mm cone trial 6278-1-121 Mod Rev 21mm + 10 mm cone trial 6278-1-123 Mod Rev 23mm + 10 mm cone trial 6278-1 -125 Mod Rev 25 mm + 10 mm cone trial 6278-1-127 Mod Rev 27mm + 10 mm cone trial 6278-1-129 Mod Rev 29mm + 10mm cone trial 6278-1-131 Mod Rev 31mm + 10 mm cone trial 6278 -1-219 Mod Rev 19mm + 20 mm cone trial 6278-1-221 Mod Rev 21mm +20 mm cone trial 6278-1-223 Mod Rev 23mm + 20 mm cone trial 6278-1-225 Mod Rev 25mm + 20 mm cone trial 6278-1-227 Mod Rev 27mm + 20 mm cone trial 6278-1-229 Mod Rev 29mm + 20 mm cone trial 6278-1-231 Mod Rev 31mm + 20 mm cone trial 6278-1-319 Mod Rev 19mm + 30 mm cone trial 6278-1-321 Mod Rev 21mm +30 mm cone trial 6278-1-323 Mod Rev 23mm + 30 mm cone trial 6278-1-325 Mod Rev 25mm + 30 mm cone trial 6278-1-327 Mod Rev 27mm + 30 mm cone trial 6278-1-329 Mod rev 29mm +30 mm cone trial 6278-1-331 Mod Rev 31mm + 30 mm cone trial 1 single unit to each non-sterile package Numerous lots numbers please see attachment.

Distribution pattern

Nationwide distribution