Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 62711

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 27, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Integra Limited

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

The MAYFIELD¿ Horseshoe Gel Pads are used in conjunction with the MAYFIELD¿ Base Units and MAYFIELD¿ Swivel Adaptors for neurosurgeries in the prone or supine positions. Those devices provide cranial stability and immobility for various craniotomies and spinal procedures. The Gel Pads are supplied to users in labeled plastic bags, 1 per bag. Product Usage: The MAYFIELD¿ Horseshoe Gel Pads are used in conjunction with the MAYFIELD¿ Base Units and MAYFIELD¿ Swivel Adaptors for neurosurgeries in the prone or supine positions. They are plastic gel filled devices that provide cushioning between the head and other metallic part of the MAYFIELD¿ system. The Left and Right Pads are mirror images.

Z-2457-2012
Recall number
Z-2457-2012
Initiated
July 27, 2012
Classification
Class II
Status
Terminated
Recalling firm
Integra Limited
Quantity
11 from Lot 0938990

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An investigation of customer complaints identified that for two product lots, Left Horseshoe Gel Pads were packaged as a Right Horseshoe Gel Pads and vice versa. None of the complaints resulted in a report of patient injuries or an increase in the time to complete the surgical procedure. A physician was consulted to assess patient risk and concluded that there is no potential harm to patients, onl

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Packaging process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An investigation of customer complaints identified that for two product lots, Left Horseshoe Gel Pads were packaged as a Right Horseshoe Gel Pads and vice versa. None of the complaints resulted in a report of patient injuries or an increase in the time to complete the surgical procedure. A physician was consulted to assess patient risk and concluded that there is no potential harm to patients, onl

Code information

The Lot No's for the affected Gel Pads are Lot 093990 of part number #440C1012 for Left Gel Pads

Distribution pattern

Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DC, GA, IA, IL, IN, NC, and NY and the countries of: Canada and Belgium.

device · product 2 of 2

The MAYFIELD¿ Horseshoe Gel Pads are used in conjunction with the MAYFIELD¿ Base Units and MAYFIELD¿ Swivel Adaptors for neurosurgeries in the prone or supine positions. Those devices provide cranial stability and immobility for various craniotomies and spinal procedures. The Gel Pads are supplied to users in labeled plastic bags, 1 per bag. Product Usage: The MAYFIELD¿ Horseshoe Gel Pads are used in conjunction with the MAYFIELD¿ Base Units and MAYFIELD¿ Swivel Adaptors for neurosurgeries in the prone or supine positions. They are plastic gel filled devices that provide cushioning between the head and other metallic part of the MAYFIELD¿ system. The Left and Right Pads are mirror images.

Z-2458-2012
Recall number
Z-2458-2012
Initiated
July 27, 2012
Classification
Class II
Status
Terminated
Recalling firm
Integra Limited
Quantity
18 from Lot 093984 for Right Gel Pads

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An investigation of customer complaints identified that for two product lots, Left Horseshoe Gel Pads were packaged as a Right Horseshoe Gel Pads and vice versa. None of the complaints resulted in a report of patient injuries or an increase in the time to complete the surgical procedure. A physician was consulted to assess patient risk and concluded that there is no potential harm to patients, onl

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Packaging process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An investigation of customer complaints identified that for two product lots, Left Horseshoe Gel Pads were packaged as a Right Horseshoe Gel Pads and vice versa. None of the complaints resulted in a report of patient injuries or an increase in the time to complete the surgical procedure. A physician was consulted to assess patient risk and concluded that there is no potential harm to patients, onl

Code information

The Lot No's for the affected Gel Pads are Lot 093984 of part number# 440C1011 for Right Gel Pads

Distribution pattern

Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DC, GA, IA, IL, IN, NC, and NY and the countries of: Canada and Belgium.