Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 62732

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 20, 2012
Product types
Device
Classifications
Class III
Statuses
Terminated
Recalling firm wording
Terumo Cardiovascular Systems Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

System 8000 Base Product Usage: The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Z-2218-2012
Recall number
Z-2218-2012
Initiated
June 20, 2012
Classification
Class III
Status
Terminated
Quantity
3

App-derived interpretation

Labeling or packaging reason.labeling_packaging · v1.0.0
mislabeling

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling mix-ups

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

During a retrospective review of complaint events, TCVS identified on 06/15/12 that three (3) Sarns" System 8000 bases were manufactured and distributed from 11/13/06-11/29/06 with an incorrect incorrect amperage rating identified on the name plate label. The mislabeling was identified in January 2007 a set up process of the automark label printer. All units were corrected in the field in Marc

Code information

catalog number: 16401 and serial number: 2001, 2002, and 2003.

Distribution pattern

US Nationwide Distribution - including the state of New York