Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 62768

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 15, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Sunquest Information Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Sunquest Laboratory Sunquest Laboratory is intended for use by professionals working in a clinical laboratory. Sunquest Laboratory is intended for use only by professionals who have received extensive training in the use of the software. Sunquest Laboratory is controlled via userdefined security levels, which allow the laboratories to manage data security and integrity. The system is marketed as an automated solution for managing laboratory processes, including: " Patient registration " Order entry and order modification " Specimen collection, verification, suitability, and distribution " Quality assurance checking " Workload recording " Billing charge capture " Standards and controls recording " Test order result inquiry and reporting Organism susceptibility and epidemiology records " Microbiology culture direct examination and observation recording " Outreach clients, specimens, and results " Client-specific reporting capabilities " Inbound client service request tracking " Historical data Clients that use the Sunquest Blood Bank system should refer to the intended use statement found in the Blood Bank System Overview Users Guide.

Z-2332-2012
Recall number
Z-2332-2012
Initiated
March 15, 2012
Classification
Class II
Status
Terminated
Quantity
126 sites

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Sunquest is recalling Sunquest Laboratory versions 6.1.0125 through 6.1.0139, 6.2.1, 6.2.2, 7.0, and 7.0.1 because when the Windows-based merge application is used and the "Event merge across patients" option is used, a receiving system (for example a HIS or CoPath system) may interpret the event merge as a full patient merge. This occurs because a trigger from the Laboratory system for the event

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sunquest is recalling Sunquest Laboratory versions 6.1.0125 through 6.1.0139, 6.2.1, 6.2.2, 7.0, and 7.0.1 because when the Windows-based merge application is used and the "Event merge across patients" option is used, a receiving system (for example a HIS or CoPath system) may interpret the event merge as a full patient merge. This occurs because a trigger from the Laboratory system for the event

Code information

Version versions 6.1.0125 through 6.1.0139, 6.2.1, 6.2.2, 7.0, and 7.0.1

Distribution pattern