Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 62769

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 08, 2012
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
I-Flow Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

ON-Q pump with ONDEMAND bolus button, ON-Q C-bloc with ONDEMAND; ON-Q C-bloc with Select a Flow and ONDEMAND Model/Reference #: 7000180, 7000181, 7000182, 7000367, 7000397, CB003, CB005, CB006, P400X1-7+5/60, P400X2-14+5/30, P400X5+5/60, and PMB01. Part Numbers: 101347100, 101347300, 101347400, 101360200, 101368400, 101368600, 101368700, 102987000, 102987200, 102987300, 103161800, 103534300, 5001438, 5001470, 5001651, 5001780, 5001808, 5001809, 5001812, 7000180, 7000181, 7000182, & 7000367. Indicated for continuous and or intermittent delivery of medication to or around surgical wound sites and or close proximity to nerves.

Z-2253-2012
Recall number
Z-2253-2012
Initiated
May 08, 2012
Classification
Class I
Status
Terminated
Recalling firm
I-Flow Corporation
Quantity
80,646 units total (65,926 units in US)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
It was determined that in a small quantity of ON-Q pumps with ONDEMAND, the bolus button may not lock in the down position when depressed and/or the orange bolus refill indicator may stay in the lowermost position. As a result, the patient may receive continuous infusion at a rate greater than expected.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

It was determined that in a small quantity of ON-Q pumps with ONDEMAND, the bolus button may not lock in the down position when depressed and/or the orange bolus refill indicator may stay in the lowermost position. As a result, the patient may receive continuous infusion at a rate greater than expected.

Code information

Products having a 10-digit Lot # that is greater than 0200521454 (i.e. 0200521455 and greater) are not impacted by this recall.

Distribution pattern