device · product 1 of 1
Orthofix ISKD Limb Lengthener. ISKD - Intramedullary Skeletal Kinetic Distractor. The ISKD is intended for limb lengthening of the tibia and femur.
- Recall number
- Z-0895-2013
- Initiated
- August 06, 2012
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Orthofix, Inc
- Quantity
- 1224
App-derived interpretation
There is a possibility that the ISKD limb lengthener may stop distracting post-operatively during treatment, which may result in premature bone consolidation (limb not achieving the desired length) leading to revision surgery to remove and/or replace the device.
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
There is a possibility that the ISKD limb lengthener may stop distracting post-operatively during treatment, which may result in premature bone consolidation (limb not achieving the desired length) leading to revision surgery to remove and/or replace the device.
Code information
All serial numbers of the following Catalog Numbers: F12-255-305, F12-345-395NS, T12-245-295, F12-255-305NS, F12-345-425, T12-255-305, F12-255-335, F12-345-425NS, T12-255-305NS, F12-255-335NS, T10-215-265, T12-255-335, F12-300-350, T10-255-305, T12-255-335NS, F12-300-350NS, T10-255-335, T12-300-350, F12-300-380, T10-300-350, T12-300-350NS, F12-300-380NS, T12-215-265, T12-300-380, F12-345-395, T12-215-265NS, and T12-300-380NS
Distribution pattern
Worldwide Distribution - USA (nationwide) Internationally to Australia, Hong Kong, Saudi Arabia, Korea, Turkey, Greece, Finland, South Africa, Spain, Singapore, Switzerland, Brazil, Germany, France, Japan, Netherlands, New Zealand, Israel, Canada and Mexico