Dimension Vista(R) Systems Vista(R) Theophylline THEO Flex(R) Regent Cartridge (K4071) The THEO method is an in vitro diagnostic test for the quantitative measurement of Theophylline in human serum and plasma on the Dimension Vista(R) system.
- Recall number
- Z-0408-2013
- Initiated
- August 09, 2012
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 959 cartons
App-derived interpretation
Firm has received complaints from customers of "Abnormal Reaction" errors occurring on calibrations, QC and patient samples. The Abnormal Reaction flag is being triggered by an internal check in the method for atypical particle reagent aggregation.
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
Firm has received complaints from customers of "Abnormal Reaction" errors occurring on calibrations, QC and patient samples. The Abnormal Reaction flag is being triggered by an internal check in the method for atypical particle reagent aggregation.
Code information
Lots 12135AE and 12171BB
Distribution pattern
USA (nationwide) including the states of AL, AZ, CA, CO, DC, DE, FL, GA, IL, IN, KY, LA, MA, MD, MI, MO, MS, MT, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and WI.