device · product 1 of 1
Siemens Mobilett Mira mobile x-ray system Product Usage: Mobile x-ray system
During regular product monitoring, firm became aware of a potential issue with the Mobilett Mira system, (serial numbers 1034, 1038, 1041, 1042, 1043, 1050, 1053, and 1054. During normal clinical operation the center screw on the driving wheel may become loose resulting in system immobilization. There is no risk of the system tipping over or collapsing on the floor, however the system may tilt t
These labels are deterministic app interpretations, not FDA categories.
During regular product monitoring, firm became aware of a potential issue with the Mobilett Mira system, (serial numbers 1034, 1038, 1041, 1042, 1043, 1050, 1053, and 1054. During normal clinical operation the center screw on the driving wheel may become loose resulting in system immobilization. There is no risk of the system tipping over or collapsing on the floor, however the system may tilt t
Code information
Model number 10273100 - serial numbers 1034, 1038, 1041, 1042, 1043, 1050, 1053, and 1054
Distribution pattern
US Nationwide Distribution including the states of: GA, IA, KS, NC and TX.