Event 63001

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 22, 2010
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Kimberly-Clark Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Kimberly-Clark MIC Safety PEG Kit - Pull Type The Kimberly-Clark MIC PEG Kits - Pull type are Percutaneous Endoscopic gastrostomy (PEG) tubes intended for a non-surgical "pull" endoscopic placement technique. The PEG tube is held in place during use with an external retention disk.

Z-2415-2012
Recall number
Z-2415-2012
Initiated
June 22, 2010
Classification
Class II
Status
Terminated
Quantity
277 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The placement guidewire diameter is too large and does not pass-through the introducer needle during the placement procedure of the PEG Feeding Tube.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The placement guidewire diameter is too large and does not pass-through the introducer needle during the placement procedure of the PEG Feeding Tube.

Code information

Lots AW916102A

Distribution pattern

Nationwide Distribution including AL, AK, AZ, CA, CO, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, ME, MA, MI, MN, MS, MO, NE, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VA, WA, and WI.

device · product 2 of 2

Kimberly-Clark MIC PEG Kit - Pull Type The Kimberly-Clark MIC PEG Kits - Pull type are Percutaneous Endoscopic gastrostomy (PEG) tubes intended for a non-surgical "pull" endoscopic placement technique. The PEG tube is held in place during use with an external retention disk.

Z-2416-2012
Recall number
Z-2416-2012
Initiated
June 22, 2010
Classification
Class II
Status
Terminated
Quantity
766 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The placement guidewire diameter is too large and does not pass-through the introducer needle during the placement procedure of the PEG Feeding Tube.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The placement guidewire diameter is too large and does not pass-through the introducer needle during the placement procedure of the PEG Feeding Tube.

Code information

Lots AW916002A, AW916102D

Distribution pattern