The Brilliance CT 16 System Product Usage: The Brilliance 16 is a Whole Body Computed Tomography X-ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles.
- Recall number
- Z-2404-2012
- Initiated
- August 28, 2012
- Classification
- Class III
- Status
- Terminated
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Quantity
- 93 units
App-derived interpretation
software version 2.3.6 is now longer and may sometimes fail to
Official device-enrichment evidence · Sourced
Software change control
Inspect official wording and provenance
Reason for recall
Philips was notified that the system logout in software version 2.3.6 is now longer and may sometimes fail to logout, requiring a forced system restart or hard shutdown with the power button.
Code information
System Code #, 728246, Serial #, 6462, 6447, 6495, 6514, 6467, 6482, 6473, 6544, 6439, 6453, 6436, 6536, 6448, 6449, 6471, 6498, 6502, 6504, 6452, 6510, 6508, 6438, 6513, 6456, 6505, 6457, 6442, 6454, 6461, 6469, 6485, 6458, 6475, 6487, 6531, 6520, 6490, 6470, 6437, 6497, 6481, 6489, 6488, 6455, 6543, 6499, 6493, 6443, 6463, 6476, 6477, 6478, 6526, 6545, 6546, 6518, 6491, 6450, 6451, 6468, 6464, 6466, 6459, 6472, 6444, 6445, 6529, 6530, 6440, 6532, 6515, 6486, 6507, 6512, 6483, 6524, 6480, 6479, 6509, 6460, 6441, 6465, 6474, 6484, 6492, 6503, 6511, 6517, 6519, 6523, 6525, 6549, 6555
Distribution pattern
Worldwide Distribution - US Nationwide including the states of: AL, CA, GA, MA, MD, NC, OH, PA, TX, and WA; and to the following countries of: Algeria, Oman, Australia, Brazil, China, Saudi Arabia, United Arab Emirates, and India.