Event 63155

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 11, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Beckman Coulter Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

AutoMate 1250 and AutoMate 2550, Part Numbers ODL25125 and ODL25255. The AutoMate 1200/2500 System Series is a semi-open, pre- and post-analytical sample processing and sorting system. The base system automates the sample sorting, decapping, and archiving processes. Handling and sorting of samples includes automatic detection of the tube type and cap type (color). Optional features include an Aliquot Module for creation of labeled secondary tubes (including the detection of the presence of adequate volume for the requested aliquots) and a Recapper Module to re-seal previously decapped tubes prior to archiving.

Z-0056-2013
Recall number
Z-0056-2013
Initiated
April 11, 2012
Classification
Class II
Status
Terminated
Recalling firm
Beckman Coulter Inc.
Quantity
468 units total (51 units in US)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The recall was initiated because Beckman Coulter has confirmed customer reports of secondary tubes being labeled with two different barcode labels on the AutoMate 2500 Family (previously referred to as OLA2500 Systems) units which can cause misidentification of a sample when software versions lower than version 4.1 is used.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling False and Misleading

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The recall was initiated because Beckman Coulter has confirmed customer reports of secondary tubes being labeled with two different barcode labels on the AutoMate 2500 Family (previously referred to as OLA2500 Systems) units which can cause misidentification of a sample when software versions lower than version 4.1 is used.

Code information

All Serial Numbers

Distribution pattern

Worldwide Distribution - USA (nationwide) and the following countries: Argentina Australia Austria Belarus Belgium China Croatia Czech Republic Denmark Finland France Germany Hungary India Ireland Israel Italy Japan Korea, Republic of Kuwait Macao Malaysia Netherlands New Zealand Norway Panama Russian Federation Slovakia Slovenia South Africa Spain Sweden Switzerland Turkey United Kingdom