device · product 1 of 1
Siemens Magnetom Skyra, Verio, Spectra and Biograph MMR Nuclear Magnetic Resonance Imaging System and Tomographic Imager
- Recall number
- Z-2464-2012
- Initiated
- September 12, 2012
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 370
App-derived interpretation
When positioning the Flex Large 4 Coil (part No. 8625761) off-center in left-right direction as it is used for example for hip and shoulder imaging and, in addition at an off-center position with respect to the head-foot direction, the coil can heat up in the area of the electronic housing (white plastic lids).
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
When positioning the Flex Large 4 Coil (part No. 8625761) off-center in left-right direction as it is used for example for hip and shoulder imaging and, in addition at an off-center position with respect to the head-foot direction, the coil can heat up in the area of the electronic housing (white plastic lids).
Code information
Model numbers: 10276755 (Verio), 10684333 (Verio Dot), 10684334 (Verio Dot upgrade), 10432915 (Skyra), 10655588 (Spectra), and 10433372 (Biograph mMR)
Distribution pattern
Nationwide Distribution - USA including Washington, DC and AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI and Puerto Rico.