Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 63235

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 08, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Parks Medical Electronics, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Devices includes a Multi Port Inflator or without Multi Port Inflator (MPI). All the devices are vascular blood flow machines. Devices are distributed with pressure cuffs of varying sizes as accessories. Affected models include *2100-SX, 3000 series, 2017, 2016, 2015, 2014, 1059, 1059-A, *1059-C, *1058-C, 4000 series, 1100 series, 1060 series, 1080 series, 234 and 246. (* = current production model) All the devices are used in the study and diagnosis of vascular disease and blockages in the upper and lower extremities.

Z-0443-2013
Recall number
Z-0443-2013
Initiated
August 08, 2012
Classification
Class II
Status
Terminated
Quantity
349 units (Domestic: 270 units - Foreign: 79 units)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Parks Medical Electronics, Inc. was made aware of two incidents where an air supply tube from their PARKS PVR unit was inadvertently connected to an IV in the patient's arm instead of into the blood pressure cuff.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Employee error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Parks Medical Electronics, Inc. was made aware of two incidents where an air supply tube from their PARKS PVR unit was inadvertently connected to an IV in the patient's arm instead of into the blood pressure cuff.

Code information

Domestic Distribution: 013053, 013104, 013107, 013124, 013154, 013182, 013193, 014203, 014237, 014248, 014249, 014252, 014253, 014254, 014269, 014276, 014310, 014335, 014339, 014361, 014366, 014376, 014380, 014381, 014382, 014384, 014397, 014401, 014406, 014416, 014424, 014429, 014431, 014443, 014444, 014448, 014454, 014458, 014530, 014532, 014533, 014540, 014542, 014545, 014547, 014551, 014552, 014552, 014553, 014598, 014601, 014604, 014608, 014616, 014654, 014657, 014659, 014660, 014661, 014663, 103077, 105X-600001024, 105X-600001025, 105X-600001026, 105X-600001027, 105X-800001019, 105X-800001022, 105X-800001023, 105X-800001024, 105X-800001026, 105X-800001027, 105X-800001027, 105X-800001028, 105X-800001029, 105X-800001030, 105X-800001031, 105X-E00001001, 105X-E00001002, 105X-E00001003, 105X-E00001004, 105X-E00001005, 105X-E00001007, 105X-E00001008, 105X-E00001009, 105X-E00001010, 105X-E00001011, 105X-E00001012, 105X-E00001013, 105X-E00001014, 105X-E00001016, 105X-E00001017, 105X-E00001018, 105X-E00001019, 105X-E00001020, 105X-F00001000, 105X-F00001001, 105X-F00001002, 105X-F00001003, 105X-F00001004, 105X-F00001005, 105X-F00001005, 105X-F00001006, 105X-F00001007, 105X-F00001008, 105X-F00001009, 105X-F00001010, 105X-F00001011, 105X-F00001012, 105X-F00001013, 105X-F00001014, 105X-F00001015, 105X-F00001016, 105X-F00001017, 105X-F00001018, 105X-F00001020, 105X-F00001021, 105X-F00001022, 105X-F00001023. There are also devices that pre-dated bar codes, as well as devices with unknown codes that the firm is unable to identify. Foreign distribution: 01301, 013043, 013044, 013045, 013046, 013072, 013076, 013175, 013181, 013183, 013188, 014215, 014386, 014392, 014393, 014456, 014535, 014550, 014551, 014561, 014562, 014563, 105X-800001025, 105X-922001017, 105X-922001018, 105X-922001019, 105X-922001021, 105X-9221020, 105X-923001001, 105X-923001002, 105X-923001003, 105X-923001004, 105X-923001005, 105X-923001006, 105X-923001007, 105X-923001007, 105X-923001009, 105X-923001010, 105X-923001011, 105X-E00001000, 105X-E00001003, 105X-E00001006, 105X-E00001015, 105X-F00001019. There are also devices that pre-dated bar codes, as well as devices with unknown codes that the firm is unable to identify.

Distribution pattern

Worldwide distribution: USA (nationwide) and countries of: Australia, Brazil, Canada, Chile, Egypt, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Korea, Mexico, Peru, Republica Argentina, Spain, Switzerland, and Venezuela.