Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 63322

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 21, 2012
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Ethicon, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

ETHICON SURGIFLO HEMOSTATIC MATRIX KIT Plus FlexTip with Thrombin distributed by ETHICON ETHICON Inc. Somerville, NJ 08876 USA Ref 2993 SURGIFLO Hemostatic Matrix, mixed with sterile saline or thrombin solution, is indicated for surgical procedures (except ophthalmic) for hemostasis, when control of capillary, venous and arteriolar bleeding by pressure, ligature, and other conventional procedures is ineffective or impractical. This is a topical absorbable hemostat consisting of a flowable gelatin matrix (paste) and a lyophilized Thrombin component. This kit also contains components necessary for preparing and applying the matrix during surgery.

Z-0292-2013
Recall number
Z-0292-2013
Initiated
August 21, 2012
Classification
Class I
Status
Terminated
Recalling firm
Ethicon, Inc.
Quantity
28,276

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
SURGIFLO Hemostatic Matrix Kit & SURGIFLO Hemostatic Matrix Kit with Thrombin have an issue within the packing process where a cut could potentially breach the double Tyvek pouch of the packaging.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Packaging

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

SURGIFLO Hemostatic Matrix Kit & SURGIFLO Hemostatic Matrix Kit with Thrombin have an issue within the packing process where a cut could potentially breach the double Tyvek pouch of the packaging.

Code information

PMA: P990004 Medical Device Listing Number: D100622 Product Code 2993 Lot/Batch Numbers: 239346 Exp Jan-13 239411 Exp Jun-13

Distribution pattern

Worldwide Distribution - USA (nationwide) and the countries of Russia, China, and the European Union.

device · product 2 of 2

ETHICON SURGIFLO HEMOSTATIC MATRIX Distributed by ETHICON ETHICON, Inc. Somerville, NJ 08876, USA SURGIFLO Hemostatic Matrix is a topical absorbable hemostat consisting of a flowable gelatin matrix (paste). This kit also contains components necessary for preparing and applying the matrix during surgery. SURGIFLO Hemostatic Matrix, mixed with sterile saline or thrombin solution, is indicated in surgical procedures (other than ophthalmic) as an adjunct to hemostasis when control of bleeding by ligature or other conventional methods is ineffective or impractical.

Z-0293-2013
Recall number
Z-0293-2013
Initiated
August 21, 2012
Classification
Class I
Status
Terminated
Recalling firm
Ethicon, Inc.
Quantity
45,200

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
SURGIFLO Hemostatic Matrix Kit & SURGIFLO Hemostatic Matrix Kit with Thrombin have an issue within the packing process where a cut could potentially breach the double Tyvek pouch of the packaging.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Packaging

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

SURGIFLO Hemostatic Matrix Kit & SURGIFLO Hemostatic Matrix Kit with Thrombin have an issue within the packing process where a cut could potentially breach the double Tyvek pouch of the packaging.

Code information

PMA number P990004 Medical Device Listing Number D100622 Product Code: 2991 Lot # 239091 Exp Mar-13 Lot # 238783 Exp Sep-13 Lot # 238635 Exp Mar-13 Lot # 237498 Exp Mar-13 Lot # 237119 Exp Mar-13 Lot # 235488 Exp Jan-13

Distribution pattern

Worldwide Distribution - USA (nationwide) and the countries of Russia, China, and the European Union.