Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 63344

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 16, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Terumo Cardiovascular Systems Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.

Z-0278-2013
Recall number
Z-0278-2013
Initiated
October 16, 2012
Classification
Class II
Status
Terminated
Quantity
19 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up

Code information

Catalog number: 809810 and serial numbers: 1041, 1083, 1084, and 1101-1116.

Distribution pattern

Nationwide Distribution including CA, HI, KS, LA, MI, MO, OR, PA, UT, and WI

device · product 2 of 2

Tested AC Control BD, 120V, X2 The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.

Z-0279-2013
Recall number
Z-0279-2013
Initiated
October 16, 2012
Classification
Class II
Status
Terminated
Quantity
15 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up

Code information

Catalog number: 817168 and serial numbers: 824POOO023, 831 P001130, 831 P001133, 831 P001136, 831 P001140, 831 POO1141 , 831 P001144, 831 P001145, 831 P001146, 831 P001149, 831 P001151, 831 P001159, 831P001161, 831P001164, and 836POOOO05.

Distribution pattern

Nationwide Distribution including CA, HI, KS, LA, MI, MO, OR, PA, UT, and WI