Event 63381

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 20, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Sorin Group USA, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

SCP Pump Control Panel, Product Code: 60-02-15, Serial Numbers: 60S09121, 60S09123-60S09129, 60S09136-60S09139. The SCP is a cardiopulmonary bypass speed control device indicated for use exclusively with the COBE Revolution pump head for speed-controlled pumping through cardiopulmonary bypass circuit for typical durations of six hours or less.

Z-0166-2013
Recall number
Z-0166-2013
Initiated
September 20, 2012
Classification
Class II
Status
Terminated
Recalling firm
Sorin Group USA, Inc.
Quantity
12 units (3 units USA and 9 units OUS)

App-derived interpretation

Potency or specification failure reason.potency_specification_failure · v1.0.0
out of specification

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Mixed-up of materials/components

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.

Code information

Product Code: 60-02-15, Serial Numbers: 60S09121, 60S09123-60S09129, 60S09136-60S09139.

Distribution pattern

Worldwide distribution: USA (nationwide) including states of : MI and NY; and countries of: Belgium, Canada, Japan, India, Russia, United Arab Emirates.

device · product 2 of 2

SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954. The SCP is a cardiopulmonary bypass speed control device indicated for use exclusively with the COBE Revolution pump head for speed-controlled pumping through cardiopulmonary bypass circuit for typical durations of six hours or less.

Z-0167-2013
Recall number
Z-0167-2013
Initiated
September 20, 2012
Classification
Class II
Status
Terminated
Recalling firm
Sorin Group USA, Inc.
Quantity
8 units (2 units USA and 6 units OUS)

App-derived interpretation

Potency or specification failure reason.potency_specification_failure · v1.0.0
out of specification

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Mixed-up of materials/components

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.

Code information

Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954.

Distribution pattern

Worldwide distribution: USA (nationwide) including states of : MI and NY; and countries of: Belgium, Canada, Japan, India, Russia, United Arab Emirates.