Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 63424

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 12, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
GE Healthcare, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

GE Discovery NM/CT670; GE Discovery NM 630 Product Usage: a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce: NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality.

Z-0121-2013
Recall number
Z-0121-2013
Initiated
October 12, 2012
Classification
Class II
Status
Terminated
Recalling firm
GE Healthcare, LLC
Quantity
210

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE Healthcare has recently become aware of a potential safety issue associated with the table pallet of your Discovery NM/CT 670 and/or Discovery NM630 system which may impact operation safety. At scan setup, during gantry rotation towards scanning position, a potential contact between the table-pallet-edge and the detector was identified. Such occurrence may damage the pallet front-edge and

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE Healthcare has recently become aware of a potential safety issue associated with the table pallet of your Discovery NM/CT 670 and/or Discovery NM630 system which may impact operation safety. At scan setup, during gantry rotation towards scanning position, a potential contact between the table-pallet-edge and the detector was identified. Such occurrence may damage the pallet front-edge and

Code information

Mfg Lot or Serial # 21058, 21002, 21013, 21017, 21020, 21029, 21039, 21041, 21049, 21066, 21075, 21079, 21080, 21081, 21082, 21119, 21140, 21144, 21145, 21147, 21150, 96BH8, 21001, 21008, 21034, 21050, 21056, 21104, 21132, 22003, 22020, 22038, 210088, 21042, 21048, 21061, 21077, 21085, 21094, 21115, 21122, 21123, 21137, 21143, 22051, , 21005, 21072, 21028, 21052, 21076, 21083, 21092, 21101, 21113, 21118, 21131, 21138, 21024, 22001, 22011, 22023, 21019, 21055, 21154, 83BH6 /000, 21007, 21047, 21009, 21021, 21022, 21033, 21035, 21037, 21063, 21064, 21067, 21070, 21073, 21086, 21095, 21121, 21128, 21134, 21141, 21146, 21152, 21153, 22014, 22016, 21003, 21006, 21010, 21027, 21036, 21043, 21046, 21051, 21084, 21108, 21116, 21120, 21124, 21133, 22015, 22028, 21093, 21158, 22021, 21031, 21129, 00ME1, 21004, 21014, 21053, 22005, 21012, 21059, 21151, 22022, 22031, 22035, 21089, 21100, 21136, 22044, 22071, 81BH0 /000, 82BH8 /000, 92BH7 /000, 93BH5 /000, 94BH3, 21016, 21040, 21060, 21103, 21125, 22037, 21038, 22008, 21068, 21126, 22009, 21026, 21011, 21018, 21025, 21071, 21090, 21109, 21110, 21130, 21142, 21160, 21096, 21107, 21091, 21099, 21000, 21023, 21030, 21032, 21044, 21045, 21054, 21062, 21065, 21069, 21074, 21105, 21111, 21114, 21127, 21149, 22002, 22004, 22007, 22010, 22012, 22017, 22018, 22024, 22025, 22026, 22027, 22029, 22030, 22032, 22033, 22034, 22036, 22039, 22040, 22041, 22042, 22043, 22045, 22046, 22047, 22048, 22049, 22050, 22052, 22053, 22054, 22056, 22057, 22059, 22064, 22067,

Distribution pattern

Worldwide distribution: USA (Nationwide) in the states of: AL, CA, FL, GA, IL, IN, KS, LA, MD, MA, MN, MS, MO, NJ, NY, NC, OH, OK, PA, SC, TN, TX, VA, WI and WV. AUSTRALIA, BELGIUM, CANADA, CHINA, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GUADELOUPE, HONG KONG, INDIA, IRELAND, ISRAEL,ITALY, JAPAN, REPUBLIC OF KOREA, NETHERLANDS,NORWAY, POLAND, RUSSIAN FEDERATION, SPAIN, SWITZERLAND, TAIWAN,THAILAND, TURKEY, UNITED KINGDOM.