Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 63586

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 19, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Trivascular, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

TriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The TriVascular Ovation Abdominal Stent Graft System is indicated in subjects diagnosed with an aneurysm in the abdominal aorta having vascular morphology suitable for endovascular repair, including: (1) Adequate iliac/femoral access compatible with vascular access techniques, devices, and/or accessories, (2) Non-aneurysmal proximal aortic neck: - with a length of at least 7 mm proximal to the aneurysm, - with an inner wall diameter of no less than 16 mm and no greater than 30 mm, and - with an aortic angle of d 60 degrees if proximal neck is e 10 mm and d 45 degrees if proximal neck is < 10 mm, (3) Non-aneurysmal distal iliac landing zone: - with a length of at least 10 mm, - with an inner wall diameter of no less than 8 mm and no greater than 20 mm

Z-0288-2013
Recall number
Z-0288-2013
Initiated
October 19, 2012
Classification
Class II
Status
Terminated
Recalling firm
Trivascular, Inc
Quantity
41 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Selected lots of the Fill Polymer Kit (European Model Number: TVFP14- B) in European distributor inventory only that potentially contain leaking stopcocks due to possible cracks in the stopcock component. TriVascular indicated that these Fill Polymer Kits are for use with the Ovation or Ovation Prime Abdominal Stent Graft Systems that may display leaks upon initiation of mix but prior to use in

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Selected lots of the Fill Polymer Kit (European Model Number: TVFP14- B) in European distributor inventory only that potentially contain leaking stopcocks due to possible cracks in the stopcock component. TriVascular indicated that these Fill Polymer Kits are for use with the Ovation or Ovation Prime Abdominal Stent Graft Systems that may display leaks upon initiation of mix but prior to use in

Code information

The following lot numbers are in distributor inventory in Europe: FF121211-04, FF010912-01, FF013112-01, FF020912-01, FF022212-03, FF022912-01, FF040412-01, FF051012-01

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of Belgium, Cyprus, Italy, Spain, Switzerland and Turkey. TriVascular stated that no purchased devices are in hospital inventory in Europe or the U.S.