Philips Digital Diagnost Software Version R2.0.2 Product Usage: This system is used for making X-ray exposures for diagnostics
- Recall number
- Z-0475-2013
- Initiated
- October 30, 2012
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Philips Healthcare Inc.
- Quantity
- 623 units
App-derived interpretation
With Digital Diagnost R2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "Right patient side" marker, although this can appear on the left patient side.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
With Digital Diagnost R2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "Right patient side" marker, although this can appear on the left patient side.
Code information
All Digital Diagnost systems with release 2.0.2 and 2.0.2 SP1 Models 712020, 712022
Distribution pattern
Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, India, Indonesia, Iraq, Italy, Japan, Latvia, Lithuania, Libya, Malta, Mexico, Netherlands, New Zealand, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, UAE and UK.