Access Immunoassay System, Access 2 Immunoassay System, Synchron LXi 725 Clinical System, UniCel DxC 600i Synchron Access Clinical System Part Numbers: Access 81600; Access 2 81600N, 386220, A69186, A25656, A12900; Access 2 (LXi) A15642; Access 2 Section, DxC 600i A25640. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The Synchron LXi and the UniCel DxC 600i are Access systems integrated with a chemistry analyzer.
- Recall number
- Z-0612-2013
- Initiated
- November 01, 2012
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 6,858 units
App-derived interpretation
It has been determined that, in some cases, pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. When these defects occur, the pulley may not turn at all, or may turn at an incorrect speed, resulting in inadequate washing or mixing of the contents of the reaction vessel.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
It has been determined that, in some cases, pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. When these defects occur, the pulley may not turn at all, or may turn at an incorrect speed, resulting in inadequate washing or mixing of the contents of the reaction vessel.
Code information
Part Numbers: Access 81600; Access 2 81600N, 386220, A69186, A25656, A12900; Access 2 (LXi) A15642; Access 2 Section, DxC 600i A25640
Distribution pattern
Worldwide Distribution-USA (nationwide) and the country of Canada.