Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 63642

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 05, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
OmniGuide, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

OmniGuide BP-ROBOTIC ( BeamPath Robotic), Single Use Cat. # BP-ROBOTIC The OmniGuide Fiber is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. It is indicated in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, gynecology, otorhinolaryngology, gastroenterology, neurosurgery, urology, and pulmonology, and can be used in open surgical procedures as well as endoscopic minimally invasive procedures in conjunction with rigid or flexible endoscopes, such as in laryngoscopy, gastroscopy, colonoscopy, laparoscopy, thoracoscopy, hysteroscopy and bronchoscopy

Z-0453-2013
Recall number
Z-0453-2013
Initiated
November 05, 2012
Classification
Class II
Status
Terminated
Recalling firm
OmniGuide, Inc.
Quantity
45 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Failure of glue joint, allowed a portion of the single use device to shift inside, may separate and fall into the patient

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Failure of glue joint, allowed a portion of the single use device to shift inside, may separate and fall into the patient

Code information

Lot Numbers: LA121003AO-P1 LA121003AQ-P1 LA121001CB-P1 LA121003AL-P1 LA121002AC-P1 LA121001CC-P1 LA121001BZ-P1 LA121001BZ-P1 LA121001AR-P1 LA121001AQ-P1 LA121001CA-P1 LA120926BA-P1 LA120926AS-P1 LA120919BG-P1 LA120920AF-P1 LA120924AG-P1 :

Distribution pattern

Nationwide Distribution including AZ, FL, MA, NY, OH, and OR.