Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 63814

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 11, 2012
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Roche Molecular Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

COBAS AmpliPrep Instrument (CAP) with AMPLILINK (AL) Software v.3.3.5 Automated sample processing instrument used with CAP/CA and CAP/CTM. Roche Molecular Systems, Inc. 1080 US Highway 202 South Branchburg, NJ 08876 USA Made in Switzerland

Z-0614-2013
Recall number
Z-0614-2013
Initiated
October 11, 2012
Classification
Class II
Status
Terminated
Quantity
7 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Certain COBAS AmpliPrep instruments using AMPILINK software v.3.3.4 had the sample clot detection flagging feature disabled during quality control release resulting in samples that may not be flagged as having clots.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Certain COBAS AmpliPrep instruments using AMPILINK software v.3.3.4 had the sample clot detection flagging feature disabled during quality control release resulting in samples that may not be flagged as having clots.

Code information

CAP 394436 to 394663

Distribution pattern