device · product 1 of 6
Sarns flexible arterial cannula, 6.7 mm (20Fr) OD with 3/8 in. connector, suture ring, 9.5 in (24 cm) long The Sarns" High Flow Aortic Arch Cannula and Sarns" Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery
- Recall number
- Z-0666-2013
- Initiated
- January 07, 2013
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Quantity
- 6830 total units for all products
App-derived interpretation
Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part.
Official device-enrichment evidence · Sourced
Process control
Inspect official wording and provenance
Reason for recall
Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part.
Code information
Lot number: 0677286 and Catalog number: 13010.
Distribution pattern
Nationwide and worldwide: AL , AR , AZ , CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , ME , MI, MO , NC , ND , NV , NY , OH , OK , PA , SD , TN , TX , UT , VA , WA , and WI. Bahrain, BRAZIL, CANADA, CHILE, Honduras, Singapore, Taiwan, Trinidad and Tobago, and UNITED ARAB EMIRATES (UAE).