Timeline bucket
December 18, 2012
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Ad-Tech Medical Instrument Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode. For temporary (<30 days) use with recording, monitoring and stimulation equipment, for the recording, monitoring and stimulation of electrical signals on the surface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.

Z-0892-2013
Recall number
Z-0892-2013
Initiated
December 18, 2012
Classification
Class I
Status
Terminated
Quantity
144 (103 US; 41 OUS)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is explanted. Recall expanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is explanted. Recall expanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes.

Code information

REF numbers (catalog numbers): EG10A-SP10N-000, EG34B-SP10X-000, EGD7A-MP12N-000, EGE4A-SP10N-000, EGE4A-SP10X-000, QG10B-SP10X-000, QG10C-SP10N-000, QG10D-SP10N-000, QG11A-SP10X-000, QG16A-SP10X-000, QG17A-SP10X-000, QG18A-SP10N-000, QG18A-SP10X-000, QG20B-SP10N-000, QG20B-SP10X-000, QG20C-SP10N-000, QG20C-SP10X-000, QG20D-SP10X-200, QG20E-SP10X-000, QG20F-SP10X-000, QG24A-SP10N-000, QG26A-SP10N-000, QG32A-SP10X-000, QG32B-SP10X-000, QG36A-SP10N-000, QG36A-SP10X-000, QG36B-SP10N-000, QG44A-SP10X-000, QG53A-SP10X-000, QG66A-SP85N-000, QG69A-SP10N-000, QG88A-SP10X-000, QG96A-SP10N-000, QG96A-SP10X-000, QG96C-SP10X-000, QG96D-SP10X-000, QGC6A-SP07N-000, QGC8A-SP10N-000, QGC8A-SP10X-000, QGC8B-SP10X-000, QGC8C-SP10X-000, QGC8D-SP10N-000, QGC8E-SP10X-000, QGC8F-SP10X-000, QS20A-SP10N-000, QS40B-SP10X-000, Lot numbers: 208140314, 208140343, 208140344, 208140353, 208140354, 208140362, 208140369, 208140374, 208140381, 208140382, 208140387, 208140390, 208140395, 208140398, 208140399, 208140414, 208140417, 208140421, 208140431, 208140443, 208140445, 208140450, 208140452, 208140453, 208140458, 019020, 208140356R, 208140357R, 208140452R, C010180001, C702201001, C703160006, C704040001, C800250002, C800350004, C801240003, C802220003, C802220005, C802220006, C805210003, C805210004, C908221002, G012111001, G012140002, G015020002, G016240001, G017250004, G017250005, G017250006, G018011001, G018011002, G018011003, G019021002, G019021003, G019150003, G112270004, G114160001, G114160002, G115080002, G115170004, G115280005, G119020004, G211220001, G608060002, G704150001, G805270001, G900211002, G903011001, G903080003, S903080004, S905280003. Recall exanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes: REF numbers (catalog numbers): CG-64BPX-F436, CMMS-22PX-F478, , CG-32BPX-F412, CGIE-98BPX-F341, CGIE-82BPN-F377, CMMS-69PN-F480, CGIE-98BPN-F341, CG-9SN-F496, CG-32BPX-F426, CG-16BPX-F445, CG-16BPX-F446, CG-256PX-F425, CGIE-39BPX-F442, CG-46BPX-F465, CG-64BPX-F419, CG-32BPX-F418. Lot numbers: 20814044, 208140447, 208140463, 208140463, 208140476, 208140463, 08140463, 208140424, 208140439, G119221001,G212040002, G216040002, 218190, 208140431, 208140450, 208140432, 208140433, 208140420, 208140437, 208140450, 208140422, 208140422, 208140422, 208140422

Distribution pattern

Worldwide Distribution - USA Nationwide including the states of CA, GA, LA, MA, MD, MI, MN, NH, PA, RI, TX, UT, WI, NY IA, WI and the countries of: Australia, CANADA, ENGLAND, JAPAN, KOREA, SPAIN, FRANCE, and SWITZERLAND.