Exact FDA event identifier

Event 64285

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 25, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
DePuy Orthopaedics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

ReClaim Reamer Extension; Product code 2975-00-500, used with Reclaim Modular Revision Hip Prosthesis System. The ReClaim Reamer Extension is as an attachment to the distal reamers to allow the surgeon to clear soft tissue while reaming the distal femoral canal.

Z-0972-2013
Recall number
Z-0972-2013
Initiated
February 25, 2013
Classification
Class II
Status
Terminated
Recalling firm
DePuy Orthopaedics, Inc.
Quantity
750

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The tabs on the reamer extension device have broken. DePuy is notifying surgeons of the issue to increase the awareness and to ask surgeons and users to look for tabs before and after surgery to reduce any potential that the tabs are left in the patient. Once the root cause is determined and a design change is implemented, DePuy will conduct a formal swap-out of the affected devices.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The tabs on the reamer extension device have broken. DePuy is notifying surgeons of the issue to increase the awareness and to ask surgeons and users to look for tabs before and after surgery to reduce any potential that the tabs are left in the patient. Once the root cause is determined and a design change is implemented, DePuy will conduct a formal swap-out of the affected devices.

Code information

Engraved Lot Number/ Label Lot Number J0111 / 578910 578913 R578910 R578912 R578913 In Kit # 2975-50-025 J0211/ In Kit # 2975-50-025 J0411/ 582631 In Kit # 2975-50-025 J0611/ In Kit # 2975-50-025 J0911/ 592125 In Kit # 2975-50-025 J1211/ 000003583 In Kit # 2975-50-025 NB12649 / NB12649 NB12650 / NB12650 NB12651 / NB12651 NB12652 / NB12652 NB12653/ NB12653 NB12654/ NB12654 NB12655/ NB12655 NB12656/ NB12656 NB23480/ NB23480 NB23481/ NB23481 NB3584 / In Kit # 2975-50-025 NB3586 / NB3586 NB3587 / NB3587 NB3588 / In Kit # 2975-50-025 NB3590 / NB3590 NB3591 / In Kit # 2975-50-025 NB5955 / NB5955 NB7894 / NB7894 In Kit # 2975-50-025 NB8016 / In Kit # 2975-50-025 NB8017 / NB8017 NB8018 / NB8018 In Kit # 2975-50-025 NB8019 / In Kit # 2975-50-025 NB8020 / NB8020 NB8021 / NB8021 NB8022 / NB8022 NB8023 /NB8023

Distribution pattern

Worldwide Distribution - USA including the states of ME, MA, PA, MD, FL, LA, IL, IA, OH, MN, TX, MT, WA, CA, HI, NV, AZ, KY, WI, CO, GA, MI, TN, NY, AR, NC and VA and the countries of Canada, Chile, Germany, Czech Republic, Israel , Norway, Poland, Slovenia, Australia, New Zealand, South Africa, Belgium, Austria, Switzerland, France, England and County Cork, Ireland.