Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 64395

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 04, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Endogastric Solutions Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

EndoGastric Solutions EsophyX2 Plus Device with Serosa Fastener and Accessories Part number C02033-01. Label reads in part "EsophyX2 2-LINK Device Qty: (1) Fastener Delivery System ***" Designed for the treatment of Gastroesophageal Reflux Disease (GERD), with erosive and non-erosive esophagitis

Z-1136-2013
Recall number
Z-1136-2013
Initiated
February 04, 2013
Classification
Class II
Status
Terminated
Recalling firm
Endogastric Solutions Inc
Quantity
190

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
EsophyX2Plus device was not validated under actual or simulated use conditions prior to distributing the product to physicians.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Release of Material/Component prior to receiving test results

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

EsophyX2Plus device was not validated under actual or simulated use conditions prior to distributing the product to physicians.

Code information

Lots 401378, 401392, 401409, and 401416 Additional lots manufactured from 11/26/12 to 12/07/12 had a stop ship issued issued on 11/21/12 and the product was placed in quarantine (it was not distributed). This represents an additional 175 units from a total of 187 units manufactured during this time period. 12 units were used for bench testing. The lot numbers of units in quarantine are 401432, 401434, 401436, and 401441.

Distribution pattern

Nationwide Distribution including NC, VA, WI, TN, KY, IN, FL, MI, IA, IL, TX, LA, AR, HI, CA, NV, OR, and WA,