Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 64436

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 15, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Medical Systems (Cleveland) Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Brilliance 64 and Ingenuity Computed Tomography X-Ray Systems The Brilliance 64 and Ingenuity CT are whole body Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Z-0950-2013
Recall number
Z-0950-2013
Initiated
February 15, 2013
Classification
Class II
Status
Terminated
Quantity
42 Units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
This field change order is being released to update software and customer release notes to the affected installed base.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

This field change order is being released to update software and customer release notes to the affected installed base.

Code information

Brilliance 64: 728231 & Ingenuity CT: 728326, 728323. System Code # 728231, Serial #: 9654, 91003, 95130, 95157, 95504 System Code # 728326, Serial #: 32000,300003,300004,300005,300010,300011,300012,300014,300015,300018,300027,300041,300053,300059,300066,300070,300071,300073,300074,300078,300079,300080,300082,300084,300086,300091,300095,300101,300102,300112,300127,300128,300129,300131,300132,310064 System Code # 728323, Serial #: 320033

Distribution pattern

Worldwide Distribution - USA including AR, AZ, IN, KS, LA, MA, MN, MS, MT, NJ, NY, NY, OH, OR, PA, TN, TX, VT, WV, WY, and worldwide including the contries of Australia, Belgium, China, France, Germany, Israel, Japan, Norway, Sweeden, and Switzerland.