device · product 1 of 1
Intraosseous Fixation System (IFS) 1.7 mm, 2.0 mm & 2.5 mm Cannulated Drill Bit. Intended to stabilize and aid in fixation of fractures.
Lot # 2399041, 97 pcs; Lot # 2399051, 143 pcs; Lot # 2399051R, 4 pcs & Lot # 2399061, 119 pcs
App-derived interpretation
Unknownreason.no_named_rule · v1.0.0
The firm was notified by a customer that there was a burr on the inside end of the drill bit resulting in the drill bit not fitting over the associated guide wire.
These labels are deterministic app interpretations, not FDA categories.
The firm was notified by a customer that there was a burr on the inside end of the drill bit resulting in the drill bit not fitting over the associated guide wire.
Code information
Part #: IFS-072-17-C, Lot #: 2399041; Part #: IFS-72-20-C, Lot #: 2399051 & Lot #: 2399051R and Part #: IFS-072-25-C, Lot #: 2399061.
Distribution pattern
Nationwide Distribution including the states of CA, CO, IL, MI, MD, NC, NM, OH, OK, PA, SD, TX, UT, VA, WA and WI.