Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 64569

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 08, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Zimmer, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Zimmer NexPosure, retractor portals of the NexPosure MIS Access System, Part Numbers or REF numbers: N1830R80-100, N1830R80-090¿, N1830R80-080¿, N1830R80-070¿, N1830R80-060, N1830R50-100, N1830R50-090, N1830R50-080, ¿ N1830R50-070, N1830R50-060, N1830L80-100, N1830L80-090, N1830L80-080, N1830L80-070, N1830L80-060, N1830L50-100, N1830L50-090, N1830L50-080, N1830L50-070, N1830L50-060, ¿ N182450-085, N182450-075, N182450-065, N182450-055, N182450-045.¿ Retractor access system that allows placement of posterior spinal fixation implants.

Z-1045-2013
Recall number
Z-1045-2013
Initiated
February 08, 2013
Classification
Class II
Status
Terminated
Recalling firm
Zimmer, Inc.
Quantity
58

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Possibility that the inner pouch may be compromised. The product is packaged in two pouches - a sterile inner pouch within an external pouch. NexPosure Retractors contained in a breached inner pouch have an increased risk of contamination (i.e.; loss of product sterility) if proper aseptic technique is not followed when transferring the inner package into the sterile field. Patient infection may o

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Possibility that the inner pouch may be compromised. The product is packaged in two pouches - a sterile inner pouch within an external pouch. NexPosure Retractors contained in a breached inner pouch have an increased risk of contamination (i.e.; loss of product sterility) if proper aseptic technique is not followed when transferring the inner package into the sterile field. Patient infection may o

Code information

Lot Number 61225572 61664689 61664689 61657007 61657015 61657015 61267745 61267745 61267750 61267750 61267753 61267753 61657016 61657016 61801814 61895685 61657017 61657018 61657018 61267756 61267756 61267759 61267759 61267760 61267760 61267761 61267761 61267761 61873348 61657019 61657019 61657031 61657031

Distribution pattern

Nationwide Distribution including the states of AZ, CO, DE, FL, ID, LMA, MI, NH, NJ, NM, PA, RI, TX, and VA.