Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 64593

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 04, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Arrow International Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Arrow/Teleflex Medical Berman Angiographic Balloon Catheter and Reverse Berman Angiographic Balloon Catheter. The Arrow Berman(TM) Angiographic Catheter is indicated for use in right heart catheterization for cardiac angiography, and measurement of the right heart and pulmonary artery pressures. It can be sued to inject radiopaque media and measure pressures in any chamber or vessel that can be entered from the right heart. The Arrow Berman(TM) Angiographic Catheter is also indicated for use in Balloon Occlusion Femoral Angiography (BOFA) of the ipsilateral leg for adult patients. The Arrow Reverse Berman(TM) Angiographic Catheter is indicated for balloon occlusion pulmonary angiography.

Z-1057-2013
Recall number
Z-1057-2013
Initiated
March 04, 2013
Classification
Class II
Status
Terminated
Recalling firm
Arrow International Inc
Quantity
99627ea

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The product specifications sheet, part of the Instructions for Use (IFU), for the specified product numbers, reference Renografin-76 contrast medium (viscosity of 8.4 centipoise), which is no longer available. Firm is in the process of updating the labeling for these products. In the meantime, when choosing an alternative contrast medium, health care providers are instructed to select a contrast

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The product specifications sheet, part of the Instructions for Use (IFU), for the specified product numbers, reference Renografin-76 contrast medium (viscosity of 8.4 centipoise), which is no longer available. Firm is in the process of updating the labeling for these products. In the meantime, when choosing an alternative contrast medium, health care providers are instructed to select a contrast

Code information

Part Numbers --AI-07035, AI-07131, AI-07135, AI-07037, AI-07132, AI-07136, AI-07130, AI-07133, AI-07137, AI-07130-J, AI-07134, AND AI-07138.

Distribution pattern

Worldwide Distribution - USA including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI and WV. and the countries of Australia, Dominican Republic, Uruguay, Ecuador, Chile, Taiwan, Pakistan, Colombia, Argentina, Korea Republic, New Zealand, Guatemala, Thailand, Peru, Bahamas, Philippines, Trinidad and Tobago, Japan, Ireland, India, Singapore, China, Canada and Mexico.