Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 64707

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 14, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Ortho-Clinical Diagnostics

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

VITROS Chemistry Products PHBR (Phenobarbital) Slides, REF 822 1384. For the quantitative measure of phenobarbital (PHBR) concentration in serum and plasma.

Z-1152-2013
Recall number
Z-1152-2013
Initiated
March 14, 2013
Classification
Class II
Status
Terminated
Quantity
Domestic: 11,164 units, Foreign: 594 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Ortho Clinical Diagnostics (OCD) received customer complaints of intermittent imprecision, outliers, or shifts in accuracy and control fluid values outside of expected intervals when using VITROS Chemistry Products PHBR Slides. OCD's investigation confirmed that positively biased values for patient samples could be obtained that may not be detected by quality control fluids.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Ortho Clinical Diagnostics (OCD) received customer complaints of intermittent imprecision, outliers, or shifts in accuracy and control fluid values outside of expected intervals when using VITROS Chemistry Products PHBR Slides. OCD's investigation confirmed that positively biased values for patient samples could be obtained that may not be detected by quality control fluids.

Code information

All Lot Numbers containing: GEN 38, Coating No. 0060, Expiration Date Range 03/01/13 through 07/01/13; GEN 39, Coating No. 0062, Expiration Date Range 07/01/13 through 09/01/13; GEN 40, Coating No. 0063, Expiration Date Range 10/01/13 through 12/01/13; GEN 41, Coating No. 0064, Expiration Date Range 12/01/13 through 02/01/14. [A Coating (CTG) is a unique identifier within the Lot Number of some of OCD's products. The term GEN and CTG refer to a specific segments of the reagent lot number. Current lots within expiry were all from Coatings 0060, 0062, 0063 or 0064.].

Distribution pattern

Worldwide Distribution-USA (nationwide) and Puerto Rico and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain, and Venezuela.