HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Versions 5.0.7 and Below. Used to measure absorbance and ion electrodes.
HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay.
These labels are deterministic app interpretations, not FDA categories.
HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay.
Code information
Model Numbers: P400ISE110EN02 and P400ISE110US02. All Serial Numbers. Versions 5.0.7 and below