device · product 1 of 1
PROFILE-V MEDTOXScan Drugs of Abuse Test System. For the rapid, qualitative detection of drugs of abuse in human urine,
- Recall number
- Z-1329-2013
- Initiated
- April 05, 2013
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Medtox Diagnostics Inc
- Quantity
- 221,625
App-derived interpretation
Review of validation information and customer complaints which indicated preliminary screen positive results for cocaine that were not positive on confirmatory testing.
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
Review of validation information and customer complaints which indicated preliminary screen positive results for cocaine that were not positive on confirmatory testing.
Code information
Part No. 604018 - Lot No. 6821, 6822, 6823, 6824, 6825, 6831, 6832, 6841, 6851, 6852, 6853, 6861, 6862, 6863, 6864, 6865, 6866, 6867, 6868, 6871, 6881, 6891, 6892, 6901, 6902, 6903, 6904, 6911, 6912, 6921, 6922, 6931, 6942; Part No. 604019 - Lot No. 6321, 6331, 6341, 6351; Part No. 604020 - Lot No. 6361, 6371; Part No. 604022 - Lot No. 6551, 6552, 6561, 6562, 6563, 6564, 6571, 6581, 6582, 6583, 6584, 6591, 6592, 6593, 6601, 6602, 6603, 6611, 6612, 6621, 6622, 6623, 6631, 6632, 6641, 6651, 6652, 6661, 6671, 6672, 6681; Part No. 604026 - Lot No. 6041, 6051; Part No. 604028 - Lot No. 6051, 6061, 6071, 6081, 6091,
Distribution pattern
Worldwide distribution: USA (nationwide) and in the country of: Canada.