VITROS Immunodiagnostic Products CK-MB Calibrators, REF 128 6293, IVD, Ortho-Clinical Diagnostics High Wycombe HP12 4DP, UK For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System for the quantitative measurement of CK-MB in human serum and plasma (EDTA or heparin).
- Recall number
- Z-1337-2013
- Initiated
- April 03, 2013
- Classification
- Class III
- Status
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics
- Quantity
- Domestic: 3019 units, Foreign: 905 units
App-derived interpretation
Internal testing demonstrated that opened-reconstituted VITROS CK-MB Calibrators stored at -20 degrees Celsius (-4 degrees Farenheit) generated invalid calibrations that were detected by positively biased quality control results.
Official device-enrichment evidence · Sourced
Process control
Inspect official wording and provenance
Reason for recall
Internal testing demonstrated that opened-reconstituted VITROS CK-MB Calibrators stored at -20 degrees Celsius (-4 degrees Farenheit) generated invalid calibrations that were detected by positively biased quality control results.
Code information
Lots 1400 (exp. 22 April 2013), 1410 (exp. 20 May 2013), 1420 (exp. 12 July 2013), 1430 (exp. 5 August 2013) , 1440 (exp. 10 September 2013), 1450 (exp. 18 November 2013), 1460 (exp. 18 November 2013), 1465 (exp. 13 November 2013), 1480 (exp. 3 January 2014)
Distribution pattern
Worldwide Distribution - Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.