Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 65048

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 16, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Alere San Diego, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Alere" hCG Combo Cassette (20/10 mIU/mL) Kit. Product Usage: The AlereTM hCG Combo (20/10 mIU/mL) test is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine, as an aid in the early detection of pregnancy. It is for healthcare professionals only.

Z-1401-2013
Recall number
Z-1401-2013
Initiated
April 16, 2013
Classification
Class II
Status
Terminated
Recalling firm
Alere San Diego, Inc.
Quantity
3975 kits

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The firm initiated recall of Alere" hCG Combo Cassette (20/10 mIU/mL) kit PN 92215, lots hCG1110133 and hCG1110135, because an investigation has indicated a portion of these Alere" hCG Combo Cassette (20/10 mIU/mL) PN 92215 kit lots were incorrectly packaged with the Alere" hCG Cassette (20 mIU/mL) PN 92210 Product Insert. The Alere" hCG Cassette (20 mIU/mL) Product Insert contains information for

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Packaging

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm initiated recall of Alere" hCG Combo Cassette (20/10 mIU/mL) kit PN 92215, lots hCG1110133 and hCG1110135, because an investigation has indicated a portion of these Alere" hCG Combo Cassette (20/10 mIU/mL) PN 92215 kit lots were incorrectly packaged with the Alere" hCG Cassette (20 mIU/mL) PN 92210 Product Insert. The Alere" hCG Cassette (20 mIU/mL) Product Insert contains information for

Code information

Lot/Unit Codes: hCG1110133 & hCG1110135 Part Number: 92215 Expiration date, or Expected shelf life: 9/30/13 & 10/31/13

Distribution pattern

United States Nationwide Distribution