Exact FDA event identifier

Event 65085

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 26, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Heart Sync, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Heart Sync C100-Physio Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible pacing/ cardioversion Defibrillation /ECG Monitoring electrodes Used in connection with a defibrillator to deliver an electrical current (shock) to the patient's heart during cardiac emergencies.

Z-1434-2013
Recall number
Z-1434-2013
Initiated
April 26, 2013
Classification
Class II
Status
Terminated
Recalling firm
Heart Sync, Inc
Quantity
17180 units for all part numbers

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
On 4/18/13 HEARTSYNC became aware of a recall from their wire/connector sub-assembly supplier which is used in the C100-Physio Electrodes. There were reports of equipment damage due to misalignment from tooling installed in August 2012. The result of this damage could render the device incapable of delivering the appropriate energy or shock to the patient.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

On 4/18/13 HEARTSYNC became aware of a recall from their wire/connector sub-assembly supplier which is used in the C100-Physio Electrodes. There were reports of equipment damage due to misalignment from tooling installed in August 2012. The result of this damage could render the device incapable of delivering the appropriate energy or shock to the patient.

Code information

Part number : C100-Physio and Lot numbers: Y100112-02, Y102212-03, Y121312-18, Y012413-02, Y022013-01, Y032113-01

Distribution pattern

Worldwide Distribution - USA including AL, GA, MI, MN, NC, ND, OH, PA, SD, VA, VT, OH, TX, MA, and Internationally to Korea.

device · product 2 of 2

Vermed, Inc Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible defibrillation/cardioversion/pacing/ECG Monitoring electrodes. Used in connection with a defibrillator to deliver an electrical current (shock) to the patient's heart during cardiac emergencies.

Z-1435-2013
Recall number
Z-1435-2013
Initiated
April 26, 2013
Classification
Class II
Status
Terminated
Recalling firm
Heart Sync, Inc
Quantity
17180 units for all part numbers

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
On 4/18/13 HEARTSYNC became aware of a recall from their wire/connector sub-assembly supplier which is used in the C100-Physio Electrodes. There were reports of equipment damage due to misalignment from tooling installed in August 2012. The result of this damage could render the device incapable of delivering the appropriate energy or shock to the patient.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

On 4/18/13 HEARTSYNC became aware of a recall from their wire/connector sub-assembly supplier which is used in the C100-Physio Electrodes. There were reports of equipment damage due to misalignment from tooling installed in August 2012. The result of this damage could render the device incapable of delivering the appropriate energy or shock to the patient.

Code information

Part Number: D304 and lot numbers:Y100112-02, Y102212-03, Y121312-18, Y012413-02, Y022013-01, Y032113-01

Distribution pattern

Worldwide Distribution - USA including AL, GA, MI, MN, NC, ND, OH, PA, SD, VA, VT, OH, TX, MA, and Internationally to Korea.