Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 65246

4 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 16, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Roche Diagnostics Operations, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

4 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 4

CoaguChek XS System (PST) The CoaguChek XS PT test strips are part of the CoaguChek XS System. The CoaguChek XS System measures blood-clotting time for people who are taking anticoagulation medications such as Coumadin¿ or warfarin.

Z-1581-2013
Recall number
Z-1581-2013
Initiated
May 16, 2013
Classification
Class II
Status
Terminated
Quantity
204,205 meters total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.

Code information

Model/Catalog/Part Number: 04837738001

Distribution pattern

Nationwide Distribution

device · product 2 of 4

CoaguChek XS Pro System The CoaguChek XS Pro System (CoaguChek XS Pro meter and CoaguChek XS PT Test strips) quantitatively determines prothrombin time ("PT"), using capillary blood or whole blood from a vein (non anticoagulated venous whole blood).

Z-1582-2013
Recall number
Z-1582-2013
Initiated
May 16, 2013
Classification
Class II
Status
Terminated
Quantity
537 meters total

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.

Code information

Model/Catalog/Part Number: 05530199160

Distribution pattern

Nationwide Distribution

device · product 3 of 4

CoaguChek XS Plus System The CoaguChek XS Plus system for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy.

Z-1583-2013
Recall number
Z-1583-2013
Initiated
May 16, 2013
Classification
Class II
Status
Terminated
Quantity
3343 meters

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.

Code information

Model/Catalog/Part Number: 05021537001

Distribution pattern

Nationwide Distribution

device · product 4 of 4

CoaguChek XS System (HCP) The CoaguChek XS System is intended for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy.

Z-1584-2013
Recall number
Z-1584-2013
Initiated
May 16, 2013
Classification
Class II
Status
Terminated
Quantity
86553 meters

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.

Code information

Model/Catalog/Part Number: 04837975001 and 04625412160

Distribution pattern

Nationwide Distribution