Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 65331

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 19, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Tecan US, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Tecan Freedom EVO with Touch Tools Suite Version 3.0 The intended use of the Touch Tools Suite is to run previously defined scripts and processes on a Freedom EVO platform (Touch Tools Suite is an optional software add-on for Freedom EVOware (the base operational software for the Freedom EVO 2 instrument.

Z-1560-2013
Recall number
Z-1560-2013
Initiated
February 19, 2013
Classification
Class II
Status
Terminated
Recalling firm
Tecan US, Inc.
Quantity
12 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
WIth TouchTools 3.0 after the execution of a Rich User Prompt in Display and Wait mode, door lock errors are ignored. If an operator presses "Continue" following a Rich User Prompt and the door has been left open, the script will progress (continue without warning). This could lead to the instrument moving unexpectedly and potentially cause an injury to the operator if they are physically inside t

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

WIth TouchTools 3.0 after the execution of a Rich User Prompt in Display and Wait mode, door lock errors are ignored. If an operator presses "Continue" following a Rich User Prompt and the door has been left open, the script will progress (continue without warning). This could lead to the instrument moving unexpectedly and potentially cause an injury to the operator if they are physically inside t

Code information

Instrument Freedom EVO 100 Base Unit: Material number: 10641100, Serial numbers: 1211001569, 1212000203 and 1212000205; Instrument Freedom EVO 100 MCA: Material number: 30020010, Serial number: 1301007643; Instrument Freedom EVO 150 Base Unit: Material number10641150, Serial number: 1301006708; Instrument Freedom EVO 150 MCA: Material number: 30020015, Serial number: 1211007442; Instrument Freedom EVO 150 MCA384: Material number: 30032015, Serial number: 1211011594 and 1109003270; Instrument Freedom EVO 200 Base Unit: Material number: 1641200, Serial number: 1106003923 and 1211011578 and Instrument Freedom EVO 200 MCA: Material number: 30020020, Serial number: 1208007288 and 1208007284.

Distribution pattern

Worldwide distribution: US (nationwide) including states of: CA, KY, NC, OH, TX, and WI; and countries of: France, Germany and Italy.