Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 65565

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 12, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Physio Control, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

The LIFEPAK 500 Automated External Defibrillator is a semi-automatic defibrillator that uses a patented Shock Advisory System. This software algorithm analyzes the patient's electrocardiographic (ECG) rhythm and indicates whether or not it detects a shockable rhythm. The LIFEPAK 500 AED is intended for use by personnel who are authorized by a physician/medical director and is intended for use in the pre-hospital and workplace and community environments.

Z-1773-2013
Recall number
Z-1773-2013
Initiated
June 12, 2013
Classification
Class II
Status
Terminated
Recalling firm
Physio Control, Inc.
Quantity
139 total (125 units in the US and 14 units outside US)

App-derived interpretation

Manufacturing or process control reason.manufacturing_process_control · v1.0.0
assembly error

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Employee error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The LIFEPAK 500 Automated External Defibrillator (AED) could be susceptible to malfunction due to an assembly error. If affected, the malfunction would be the result of internal shorting on a printed circuit board assembly and could render the device inoperable.

Code information

PART NUMBERS: 3011790-001568, 3011790-001574, 3011790-001627, 3011790-001634, and 3011790-001650. SERIAL NUMBERS OUTSIDE the US: 41232265, 41232267, 41232269, 41232270, 41232271, 41232272, 41238207, 41238208, 41301145, 41301370, 41301371, 41301372, 41301373, and 41301374. SERIAL NUMBERS IN THE US: 41134905, 41134906, 41134907, 41135982, 41135983, 41135984, 41135991, 41135992, 41135993, 41135994, 41135995, 41135996, 41135997, 41135998, 41135999, 41136000, 41136001, 41136002, 41136003, 41136004, 41136005, 41136006, 41136007, 41136008, 41136009, 41136010, 41136011, 41136012, 41136013, 41136014, 41136015, 41136016, 41136017, 41136018, 41136019, 41136020, 41136021, 41136022, 41136023, 41136024, 41136025, 41136026, 41136027, 41136028, 41136029, 41136030, 41136031, 41136033, 41136034, 41136035, 41136036, 41136037, 41136038, 41136039, 41136040, 41136060, 41136061, 41136062, 41136063, 41136064, 41136065, 41136066, 41136067, 41136068, 41136069, 41136070, 41136071, 41136072, 41136073, 41136074, 41136075, 41136076, 41136077, 41136078, 41136079, 41136092, 41136093, 41136094, 41142873, 41145359, 41145360, 41145361, 41145362, 41145363, 41145364, 41145365, 41145366, 41145367, 41145368, 41145369, 41145370, 41145371, 41145372, 41145373, 41145374, 41145375, 41145376, 41145377, 41145378, 41164115, 41232002, 41232003, 41232004, 41232005, 41232249, 41232250, 41232251, 41232252, 41232253, 41232254, 41232255, 41232256, 41232257, 41232258, 41232259, 41232260, 41232554, 41239182, 41239183, 41239184, 41239185, 41239186, 41239187, 41239188, and 41239189.

Distribution pattern

Worldwide distribution: US (nationwide) including states of: IL, MA, NC, ND, NY, PA, RI, TX, WA, and WI; and country of: Canada.