device · product 1 of 1
syngo Dynamics Information System (sDIS) versions 9.5 and VA10A. The syngo¿ Dynamics system is intended for the acceptance, transfer, display, storage, archive, and manipulation of digital medical images, including quantification and reportgeneration. syngo¿ Dynamics is not intended to be used for reading mammography images.
- Recall number
- Z-1859-2013
- Initiated
- June 10, 2013
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 9
App-derived interpretation
Siemens initiate this recall due to a potential issue when using the sDIS matching rules of the syngo Dynamics Information System version 9.5 or VA10A with modality device configured to end study at association close. Images or DICOM SR objects sent from a modality device to syngo Dynamics may not be saved when the modality is configured at syngo Dynamics to end study at association close, and whe
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
Siemens initiate this recall due to a potential issue when using the sDIS matching rules of the syngo Dynamics Information System version 9.5 or VA10A with modality device configured to end study at association close. Images or DICOM SR objects sent from a modality device to syngo Dynamics may not be saved when the modality is configured at syngo Dynamics to end study at association close, and whe
Code information
Model Number: 10091637, 10091805, with serial numbers 85164, 85289, 85296, 85297, 85149, 85116, 85177, 85051, and 85278.
Distribution pattern
US Distribution only including the states of PA, OH, UT, CA, NY, MO, and TX.