Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 65617

10 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 15, 2013
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Nova Biomedical Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

10 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 10

Nova Max Glucose Test Strips (50 Count Retail) Catalog Number: 43437 Product Usage: The Nova Max Glucose Test Strips are for use with the Nova Max Family of Glucose Monitors (Nova Max, Nova Max Plus, Nova Max Link, BD Logic and Paradigm Link). These monitors are intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. They are intended for use by people with diabetes mellitus in the home and by healthcare professionals in clinical settings as an aid to monitor the effectiveness of diabetes control

Z-1897-2013
Recall number
Z-1897-2013
Initiated
July 15, 2013
Classification
Class I
Status
Terminated
Quantity
22,395,802 test strips

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of false abnormally high glucose readings from test strips

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of false abnormally high glucose readings from test strips

Code information

Lot #/Exp.Date: 1010211346 12/31/2013 1020212032 02/28/2014 1020212100 04/30/2014 1020212101 04/30/2014 1020212153 06/30/2014 1020212207 07/31/2014 1020212291 10/31/2014 1020213074 03/31/2015 1020213109 04/30/2015 1020411347 12/31/2013 1020412255 09/30/2014 1020412341 12/31/2014 1020512087 03/31/2014 1020611348 12/31/2013 1020712206 07/31/2014 1020811350 12/31/2013

Distribution pattern

Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua

device · product 2 of 10

Nova Max Glucose Test Strips (50 Count DME) Catalog Number: 43523 Product Usage: The Nova Max Glucose Test Strips are for use with the Nova Max Family of Glucose Monitors (Nova Max, Nova Max Plus, Nova Max Link, BD Logic and Paradigm Link). These monitors are intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. They are intended for use by people with diabetes mellitus in the home and by healthcare professionals in clinical settings as an aid to monitor the effectiveness of diabetes control

Z-1898-2013
Recall number
Z-1898-2013
Initiated
July 15, 2013
Classification
Class I
Status
Terminated
Quantity
32,452,700 test strips

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of false abnormally high glucose readings from test strips

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of false abnormally high glucose readings from test strips

Code information

Lot #/Exp.Date: 1010312046 10/31/2013 1020212032 02/28/2014 1020212100 04/30/2014 1020212101 04/30/2014 1020212153 06/30/2014 1020212207 07/31/2014 1020212291 10/31/2014 1020213074 03/31/2015 1020213109 04/30/2015 1020412255 09/30/2014 1020412341 12/31/2014 1020512087 03/31/2014 1020712206 07/31/2014 1020811350 12/31/2013 1020912292 10/31/2014

Distribution pattern

Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua

device · product 3 of 10

Nova Max Glucose Test Strips (10 Count Boxed) Catalog Number: 43590 Product Usage: The Nova Max Glucose Test Strips are for use with the Nova Max Family of Glucose Monitors (Nova Max, Nova Max Plus, Nova Max Link, BD Logic and Paradigm Link). These monitors are intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. They are intended for use by people with diabetes mellitus in the home and by healthcare professionals in clinical settings as an aid to monitor the effectiveness of diabetes control

Z-1899-2013
Recall number
Z-1899-2013
Initiated
July 15, 2013
Classification
Class I
Status
Terminated
Quantity
721,810 test strips

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of false abnormally high glucose readings from test strips

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of false abnormally high glucose readings from test strips

Code information

Lot #/Exp.Date: 1020211355 12/31/2013 1020212032 02/28/2014 1020212207 07/31/2014 1020212291 10/31/2014 1020213043 02/28/2015 1020213109 04/30/2015 1020412255 09/30/2014 1020412341 12/31/2014 1020512087 03/31/2014 1020811350 12/31/2013

Distribution pattern

Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua

device · product 4 of 10

Nova Max Glucose Test Strips (100 Count Retail Canada)Catalog Number: 43955 Product Usage: The Nova Max Glucose Test Strips are for use with the Nova Max Family of Glucose Monitors (Nova Max, Nova Max Plus, Nova Max Link, BD Logic and Paradigm Link). These monitors are intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. They are intended for use by people with diabetes mellitus in the home and by healthcare professionals in clinical settings as an aid to monitor the effectiveness of diabetes control

Z-1900-2013
Recall number
Z-1900-2013
Initiated
July 15, 2013
Classification
Class I
Status
Terminated
Quantity
1,564,500 test strips

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of false abnormally high glucose readings from test strips

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of false abnormally high glucose readings from test strips

Code information

Lot #/Exp.Date: 1020212032 02/28/2014 1020212100 04/30/2014 1020212153 06/30/2014 1020212207 07/31/2014 1020213074 03/31/2015 1020412341 12/31/2014 1020912292 10/31/2014

Distribution pattern

Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua

device · product 5 of 10

Nova Max Glucose Test Strips (10 Count Vials) Catalog Number: 43604 These strips are kitted into Nova Max Glucose Test Strip System Kits 43435 and 45507 Product Usage: The Nova Max Glucose Test Strips are for use with the Nova Max Family of Glucose Monitors (Nova Max, Nova Max Plus, Nova Max Link, BD Logic and Paradigm Link). These monitors are intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. They are intended for use by people with diabetes mellitus in the home and by healthcare professionals in clinical settings as an aid to monitor the effectiveness of diabetes control

Z-1901-2013
Recall number
Z-1901-2013
Initiated
July 15, 2013
Classification
Class I
Status
Terminated
Quantity
*Kitted into System Kits 43435 and 45507

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of false abnormally high glucose readings from test strips

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of false abnormally high glucose readings from test strips

Code information

Lot #/Exp.Date: 1020212032 02/28/2014 1020212153 06/30/2014 1020212207 07/31/2014 1020212291 10/31/2014 1020213074 03/31/2015 1020213109 04/30/2015 1020412255 09/30/2014 1020412341 12/31/2014 1020512087 03/31/2014 1020811350 12/31/2013 102121304 02/28/2015

Distribution pattern

Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua

device · product 6 of 10

Nova Max Glucose Test Strips (50 Count International Catalog Number: 45512 Product usage: The Nova Max Glucose Test Strips are for use with the Nova Max Family of Glucose Monitors (Nova Max, Nova Max Plus, Nova Max Link, BD Logic and Paradigm Link). These monitors are intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. They are intended for use by people with diabetes mellitus in the home and by healthcare professionals in clinical settings as an aid to monitor the effectiveness of diabetes control

Z-1902-2013
Recall number
Z-1902-2013
Initiated
July 15, 2013
Classification
Class I
Status
Terminated
Quantity
366,140 test strips

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of false abnormally high glucose readings from test strips

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of false abnormally high glucose readings from test strips

Code information

Lot #/Exp.Date: 1020112086 03/31/2014 1020212032 02/28/2014 1020213074 03/31/2015 1020213109 04/30/2015

Distribution pattern

Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua

device · product 7 of 10

Nova Max Plus Monitor System (Kit w/10 Count Vials) Catalog Number: 43435 Product usage: The Nova Max Glucose Test Strips are for use with the Nova Max Family of Glucose Monitors (Nova Max, Nova Max Plus, Nova Max Link, BD Logic and Paradigm Link). These monitors are intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. They are intended for use by people with diabetes mellitus in the home and by healthcare professionals in clinical settings as an aid to monitor the effectiveness of diabetes control

Z-1903-2013
Recall number
Z-1903-2013
Initiated
July 15, 2013
Classification
Class I
Status
Terminated
Quantity
1,316,420 kits

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of false abnormally high glucose readings from test strips

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of false abnormally high glucose readings from test strips

Code information

Kit Lot/Exp: 1010112004 10/31/2013 1010112038 10/31/2013 1010112109 02/28/2014 1010112121 02/28/2014 1010112159 02/28/2014 1010112228 05/31/2014 1010112289 07/31/2014 1010113016 09/30/2014 1010113052 12/31/2014 1010113134 12/31/2014 1010113177 03/31/2015 1010212046 10/31/2013 1010212109 02/28/2014 1010212121 12/31/2013 1010212143 02/28/2014 1010212206 03/31/2014 1010212228 03/31/2014 1010212289 07/31/2014 1010212354 09/30/2014 1010213016 09/30/2014 1010213052 12/31/2014 1010213134 09/30/2014 1010312016 10/31/2013 1010312025 10/31/2013 1010312046 10/31/2013 1010312095 02/28/2014 1010312121 03/31/2014 1010312171 03/31/2014 1010312180 03/31/2014 1010312207 03/31/2014 1010312228 05/31/2014 1010312262 06/30/2014 1010312289 07/31/2014 1010312312 09/30/2014 1010312354 09/30/2014 1010313016 09/30/2014 1010313052 12/31/2014 1010313101 12/31/2014 1010313128 01/31/2015 1010313134 01/31/2015 1010411355 10/31/2013 1010412016 10/31/2013 1010412046 10/31/2013 1010412066 10/31/2013 1010412093 12/31/2013 1010412095 10/31/2013 1010412121 03/31/2014 1010412262 03/31/2014 1010412312 09/30/2014 1010413101 01/31/2015 1010413134 01/31/2015 1010413171 04/30/2015 1010512016 10/31/2013 1010512095 12/31/2013 1010512157 03/31/2014 1010512212 03/31/2014 1010512262 07/31/2014 1010512269 07/31/2014 1010512312 09/30/2014 1010513085 12/31/2014 1010513101 01/31/2015 1010513171 04/30/2015 1010612095 02/28/2014 1010612157 03/31/2014 1010612262 07/31/2014 1010612312 07/31/2014 1010612339 09/30/2014 1010613085 10/31/2014 1010613171 04/30/2015 1010712157 03/31/2014 1010712262 07/31/2014 1010713079 12/31/2014 1010713085 12/31/2014 1010813085 12/31/2014 1010912254 05/31/2014 1011113007 09/30/2014

Distribution pattern

Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua

device · product 8 of 10

Nova Max Glucose Test Strips (10 Count Canada)Catalog Number: 43877 Product The Nova Max Glucose Test Strips are for use with the Nova Max Family of Glucose Monitors (Nova Max, Nova Max Plus, Nova Max Link, BD Logic and Paradigm Link). These monitors are intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. They are intended for use by people with diabetes mellitus in the home and by healthcare professionals in clinical settings as an aid to monitor the effectiveness of diabetes control

Z-1904-2013
Recall number
Z-1904-2013
Initiated
July 15, 2013
Classification
Class I
Status
Terminated
Quantity
9,200

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of false abnormally high glucose readings from test strips

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of false abnormally high glucose readings from test strips

Code information

1020211350 12/31/2013

Distribution pattern

Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua

device · product 9 of 10

Nova Max Plus Monitor System (Kit w/10 Count Vials Canada) Catalog Number: 43861 Product usage: The Nova Max Glucose Test Strips are for use with the Nova Max Family of Glucose Monitors (Nova Max, Nova Max Plus, Nova Max Link, BD Logic and Paradigm Link). These monitors are intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. They are intended for use by people with diabetes mellitus in the home and by healthcare professionals in clinical settings as an aid to monitor the effectiveness of diabetes control

Z-1905-2013
Recall number
Z-1905-2013
Initiated
July 15, 2013
Classification
Class I
Status
Terminated
Quantity
28,240 kits

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of false abnormally high glucose readings from test strips

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of false abnormally high glucose readings from test strips

Code information

1010112073 10/31/2013 1010113002 03/31/2014 1010113008 03/31/2014 1010312097* 10/31/2013 1010512080** 10/31/2013

Distribution pattern

Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua

device · product 10 of 10

Nova Max Plus Monitor System (Kit w/10 Count Vials International) Catalog Number: 45507 Product Usage: The Nova Max Glucose Test Strips are for use with the Nova Max Family of Glucose Monitors (Nova Max, Nova Max Plus, Nova Max Link, BD Logic and Paradigm Link). These monitors are intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. They are intended for use by people with diabetes mellitus in the home and by healthcare professionals in clinical settings as an aid to monitor the effectiveness of diabetes control

Z-1906-2013
Recall number
Z-1906-2013
Initiated
July 15, 2013
Classification
Class I
Status
Terminated
Quantity
13,600 kits

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Reports of false abnormally high glucose readings from test strips

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Reports of false abnormally high glucose readings from test strips

Code information

1010112059 12/31/2013 1010213171 04/30/2015 1010312066 12/31/2013 1010712339 10/31/2014

Distribution pattern

Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua